vPad-Rugged2

vPad‐Rugged 2 is much more than an Electrical Safety Analyzer ‐ it’s the heart of your Preventive Maintenance process. Using one of our CMMS Interface Apps, work orders can be downloaded at the start of the day. Automated Preventive Maintenance routines ensure that PMs are consistently carried out per OEM requirements. Completed PM records are uploaded to your CMMS to automate closure of open work orders

Product Details

Key Features

■Tests to NFPA 99, AAMI‐ES1, IEC60601 and IEC62353 standards
■ Field upgradeable by installation of optional Apps
■ Built‐in wireless communication enables transfer of data from remote sites to corporate CMMS systems
■ Can be configured to retrieve test data from non‐Datrend Instruments

Screen Shot 2024 01 24 At 12.42.07 PM

Electrical Safety Testing Standards

  • NFPA 99
  • IEC 62353
  • IEC 60601
  • AAMI‐ES1

Safety Tests

  • Mains Voltage
  • External (point‐to‐point) Voltage
  • External (point‐to‐point) Micropotential
  • Equipment Load Current
  • Equipment Power Consumption
  • Protective Earth Resistance
  • External (point‐to‐point) Resistance
  • Equipment Leakage Current
  • Applied Part Leakage
  • Insulation Resistance

Instrument Specifications

Voltage Measurement

  • Mains voltage
    • Range…………………………………….. 90 to 264 V rms
    • Accuracy …………………………………±(2 % of reading + 0.2 V)
  • External (point‐to‐point) voltage
    • Range…………………………………….. 0 to 300 V rms
    • Accuracy …………………………………±(1 % FS + 0.2 V)
  • External (point‐to‐point) micropotential
    • Ranges
      • 0 to 199.9 mV rms
      • 200 to 1,999 mV rms
      • 2,000 to 19,999 mV rms
    • Accuracy …………………………………±(1 % of reading + 1 mV)

Load Current Measurement

  • Ranges
    • 0 to 1.999 A ac rms
    • 2.00 to 20 A ac rms
  • Accuracy ……………………………….. ±(2 % reading + 0.2A)
  • Duty cycle
    • 0 A to 10 A, continuous
    • 10 A to 15 A, 7 min. on/3 min. off
    • 15A to 20 A, 5 min. on/5 min. off

Power Measurement

  • Range ……………………………………. 0 to 2400 watts
  • Accuracy …………………………………±(5 % reading + 5W)

Earth Resistance Measurement

  • Modes ……………………………………Four‐terminal, fully isolated
  • Test Current ……………………………1A pulsed, 0.2A rms
  • Range …………………………………… 0.000 to 2.000 Ω
  • Accuracy ………………………………..±(1 % of reading + 0.02 Ω)

Insulation Resistance Measurement

  • Ranges
    • Test Voltage: 500V and 250V
    • 0.5 to 4.9 MΩ
    • 5 to 49 MΩ
    • 50 to 999.9 MΩ
    • Test Voltage: 100V and 50V
    • 0.1 to 0.9 MΩ
    • 1 to 9.9 MΩ
    • 10 to 999.9 MΩ
  • Range Selection………………………. Automatic
  • Accuracy
    • Test Voltage: 500V and 250V
    • 5 MΩ Range……………………….. ±(1 % of reading + 0.1 MΩ)
    • 50 MΩ Range……………………… ±(2 % of reading + 0.2 MΩ)
    • 50 to 500 MΩ……………………… ±(5 % of reading + 0.2 MΩ)
    • Test Voltage: 100V and 50V
    • 1 MΩ Range……………………….. ±(1 % of reading + 0.1 MΩ)
    • 10 MΩ Range……………………… ±(2 % of reading + 0.2 MΩ)
    • 10 to 100 MΩ……………………….. ±(5 % of reading + 0.2 MΩ)
  • Test Voltage
    • Selections…………………………….500V, 250V, 100V or 50V
    • Accuracy…………………………….. ±5 % for 0 to 1 mA load
    • Maximum load capacitance….. 1 μF

Equipment and Applied Part Leakage Tests (vPad‐623 app)

  • Measurement…………………………. True RMS
  • Method………………………………….. Direct, Alternative or Differential method
  • Patient Load…………………………….Per IEC 62353
  • Functional Group Selections…….. AP1; AP1 and AP2; AP3 and AP4; AP1 to AP3; AP1 to AP4; AP5 to AP10; AP1 to AP10

Chassis and Lead Leakage Tests (vPad‐ES app)

  • Measurement
    • AC+DC (True RMS)
    • AC only
    • DC only
  • Patient Load Selections……………. IEC 60601 and AAMI‐ES1

Leakage Current Measurement

  • Crest Factor…………………………….. <= 3
  • Ranges
    • Direct and Alternative Methods
      • 0.0 to 199.9 μA
      • 200 to 1,999 μA
      • 2,000 to 19,999 μA
    • Differential Method
      • 50.0 to 199.0 μA
      • 200 to 1,999 μA
      • 2,000 to 19,999 μA
  • Accuracy
    • Direct and Alternative Methods
      • DC to 1 kHz…………………….. ±(1 % of reading + 1 μA)
      • 1 to 100 kHz…………………… ±(2 % of reading + 1 μA)
      • 100 kHz to 1 MHz……………. ±}(5 % of reading + 1 μA)
    • Differential Method
      • 50.0 to 20,000 μA……………. ±(5 % of reading + 20 μA)
    • Isolation test voltage………………. 100 % ±5 % of AC supply

ECG Simulator

  • Output
    • Amplitude……………………………1 mV QRS into Lead II
    • Impedance…………………………..500 ohms
  • Accuracy
    • Frequency…………………………… ±1 %
    • Amplitude……………………………±2 %
  • Waveforms
    • ECG Complex………………………..30, 60, 120, 180, 240 and 300 BPM
    • Square wave……………………….. 0.125 Hz, 2 Hz and 1 kHz
    • Pulse wave…………………………..63 msec, 30PPM and 60PPM
    • Triangle wave……………………….2 Hz
    • Sine wave…………………………….0.5, 10, 40, 50, 60 and 100 Hz
    • CMRR tests…………………………. SQR 2Hz & 1 kHz, PUL 4 sec,
    • SIN 0.5, 50 & 60Hz
    • Arrhythmias…………………………VFIB, AFIB, SVT, VTACH, PVC and ASYS

User Interface

  • Display…………………………………….10.1″ colour LCD (1280 x 800)
  • User Controls………………………….. Capacitive touchscreen
  • Wired Connectivity
    • Micro USB 2.0 Type B
    • XBUS Interface (RJ11‐6)
  • Wireless Connectivity
    • 802.11 b/g/n
    • Bluetooth 2.1+ EDR
  • Memory Internal………………………16GB Maximum †
  • Memory Expansion…………………..16GB to 32GB Micro SD Card (optional)
  • Modes of Operation
    • Manual (standard)
    • Automatic (accessory option)

Dimensions

  • Base unit………………………………… 14.3 x 11.1 x 4.7 in. (36 x 28 x 12 cm)
  • Tablet PC………………………………… 10.5 x 6.7 x 0.5 in. (27 x 17 x 1.3 cm (detachable from Base Unit)

Weight

9.0 lb (4.1 kg)

All specifications subject to change without notice.