vPad-RF

vPad-RF – the smallest full-featured analyzer for testing Electrosurgical Units (ESUs). Built on the Vision Pad platform, vPad-RF has everything you have come to expect from the technological leader in biomedical test equipment.

Product Details

Key Features

■Digital signal processor (DSP) enables accurate testing of all ESU makes and models, continuous and pulsed-waveforms
■ Integration of ESU testing with HF digital storage oscilloscope and spectrum analyzer functions provides three pieces of test equipment in one
■ Multi-pulse analysis averages measurements from multiple RF pulses, reducing errors caused by analyzing pulses in isolation
■ High powered RF load provides settings required for all IEC 60601-2-2 tests, including Covidien ForceTriad maintenance procedures
■ Wide measurement range - up to 8.5A
■ Pre-Configured automated PM routines for most major manufacturers
■ Works with vPad safety analyzers and automation apps.

vPad-RF

vPad-RFTM – Technical Specifications

HF Power (RMS):

  • Range: 0 – 999.9 W (as derived from HF current)
  • Accuracy: ±(3% of reading + óW)

HF Current (RMS):

  • Range: 0 – 8.5A
  • Accuracy: ±(1% of reading + 5mA)

Voltage (RMS):

  • Range: 0-999.9 V (as derived from HF current)
  • Accuracy: ±(1% of reading + 1V)

Voltage (Peak):

  • Range: 0 – 9999 V
  • Accuracy: ±(1% of reading + 10V)

Crest Factor: 1.0 – 999.9 (as derived from peak/RMS ratio)

Instrument RF Bandwidth (-3dB): 250 Hz – 12 MHz

Isolation, PE to MD: 10 kV

HF Load:

Range

  • Internal: Variable 0 – 5115 ohms
  • Internal + External: Max 30,000 ohms
  • Resolution 5 ohms
  • Accuracy ±1% +0.5, -0 ohms
  • Power, 50% duty cycle
    • 200W @ 5 ohms
    • 400W @ 10 ohms
    • 600W @ 20 ohms
    • 600W @ 40 ohms
    • 600W @ 80 ohms
    • 600W @ 160 ohms
    • 600W @ 320 ohms
    • 600W @ 640 ohms
    • 530W @ 1280 ohms
    • 260W @ 2560 ohms

RF Leakage Load:

  • Resistance 2 x 200 ohms
  • Accuracy ±1%
  • Power, 100% duty 16W

REM/CQM Test Resistance:

  • Range 0 – 1023 ohms
  • Resolution 1 ohm
  • Accuracy ±1% +1, -0 ohms

Multi-pulse Averaged Results: On-time, off-time & duty cycle

Operational Duty Cycle:

Up to 100% in manually activated tests, 50% in automated footswitch activation, from 1 to 20 seconds

HF Digital Oscilloscope:

  • Vertical Ranges: 0.125, 0.25, 0.75, 1.5, or 3A / division
  • Vertical Range Selection: Manual or Automatic
  • Horizontal Ranges: 1.25, 2.5, 5, 10, 20 or 40 μsec / division
  • Trigger Level: Adjustable, +/- edge

HF Spectrum Analyzer:

  • Vertical Ranges Normalized dB or linear
  • Horizontal Ranges 62.5, 125, 250, 500, 1000, 2000 kHz / division
  • Frequency Resolution 1.22, 2.44, 4.88, 9.77, 19.5, 39.1 kHz

User Interface:

Android 10” Tablet providing Graphical User Interface, WiFi & Bluetooth communication, 12 Gb memory (or more – optional)

Power:

  • 120 Volt power outlet 90 to 132 V ac rms, 47 to 63 Hz, 0.6 A max.
  • 230 Volt power outlet 180 to 264 V ac rms, 47 to 63 Hz, 0.4 A max.

Size: 8.5 x 11 x 13 in.

Weight: 18 lb.

Environment: 15°C to 40°C, 10% to 90% RH, Indoor Use Only, Category II

All specifications subject to change without notice.